EU Manufacturer, Swiss Customer: Do You Need a Swiss Importer?

Gerd Eichhorn · November 2026

An EU manufacturer that delivers directly to Swiss hospitals and practices does not need a Swiss importer, as long as the customer uses the devices itself. Whether it needs a Swiss authorized representative (CH-REP) depends on one term: placing on the market.

Since May 2021, Switzerland and the EU treat each other as third countries for medical devices. Every EU manufacturer selling into Switzerland therefore asks the same question: do we need a CH-REP, a Swiss importer, or both? Many supply chains are simple. The manufacturer ships from its EU site straight to a Swiss hospital, with no Swiss distributor in between. This article explains who plays which role in that case, and where the open questions are.

Placing on the market: the term that decides

In Switzerland, placing on the market means the first making available of a device on the Swiss market [MedDO, SR 812.213, Art. 4(1)(b)]. Making available means any transfer of a device for distribution, consumption or use on the Swiss market in the course of a commercial activity, whether paid or free of charge [MedDO, SR 812.213, Art. 4(1)(a)]. Swissmedic adds that the transfer does not require a physical handover, and that the concept applies to each individual device, not to a device type (Swissmedic MU600_00_016, version 8.0, section 3).

The economic operator roles follow from this term:

CH-REP: a manufacturer not established in Switzerland may only place devices on the market if it has mandated a CH-REP [MedDO, SR 812.213, Art. 51(1)].

Importer: any person established in Switzerland who places a device from abroad on the Swiss market [MedDO, SR 812.213, Art. 4(1)(h)]. The role is not designated; it follows from the activity.

No placing on the market, no obligation for either role.

The term is not as settled as it looks. In 2022, a legal opinion commissioned by Swiss Medtech criticized Swissmedic’s view on when imported stock counts as placed on the market and called it inconsistent with the EU reading (Prager Dreifuss, April 22, 2022). The opinion notes that only a Swiss court ruling under the current ordinance will create legal certainty.

Direct delivery: why there is no Swiss importer

When a Swiss hospital or practice obtains a CE-marked device from abroad and uses it itself, the device is not placed on the market in Switzerland. The institution does not become an importer and is not subject to the importer’s checking, registration and documentation obligations (Swissmedic MU600_00_006, version 6.1, section 7). Instead, the professional who directly uses a device from abroad is responsible for its conformity [MedDO, SR 812.213, Art. 70(1)].

Swissmedic confirms this also applies when devices are moved within the same legal entity, for example central purchasing for several departments, and when an implant stays with the patient. It changes as soon as the institution passes devices on to another legal entity: then it places them on the market and becomes the importer.

Direct delivery is about the contract, not the truck. What counts is who acquires the device, not who physically receives it. Swissmedic defines making available as an offer or agreement between two or more persons on the transfer of ownership, possession or other rights, which does not necessarily require a physical handover. It adds that a professional applying a device, for example an implant, does not make it available on the market (MU600_00_016, section 3). MDSS illustrates the same point for the EU: the manufacturer may ship straight to the hospital while ownership passes through an importer and a distributor, who then remain importer and distributor (MDSS, Importer: Myths vs Facts, December 2024, slides 24 to 27). Applied to Switzerland: if the hospital is invoiced by a Swiss company, that company is the importer, even if the parcel comes straight from the EU. Only when the hospital buys directly from the EU manufacturer is there no importer.

Supply chainPlaced on the Swiss market?Swiss importerCH-REP
EU manufacturer → Swiss hospital, used by the hospital itselfNoNoneNot compulsory, recommended
EU manufacturer → Swiss hospital or purchasing organization → another Swiss legal entityYesThe hospital or purchasing organizationRequired
EU manufacturer → Swiss distributor → hospitalYesThe Swiss distributorRequired
EU manufacturer offers devices to Swiss users online or by other distance sellingContested (see below)NoneRequired in Swissmedic’s reading

The first three rows follow Swissmedic’s procurement information sheet. The last row is the subject of the next section.

Distance selling: a contested provision

The “no placing on the market” reading has a limit, and it is disputed where it lies. Devices offered to users in Switzerland online or by another form of distance selling are deemed to be made available on the market [MedDO, SR 812.213, Art. 7(1bis)]. Two readings exist:

QuestionSwissmedicIsler, LSR 1/2024
Is a distance sale by a foreign manufacturer a placing on the Swiss market?Yes: making available exists, and the first making available is the placing on the market (MU600_00_006, footnote 29)No: the deemed making available is a legal fiction, not a placing on the market
CH-REP required?Yes [MedDO, Art. 51(1)]No, because the CH-REP duty is triggered only by placing on the market
Swiss importer?None established in SwitzerlandNone; the ordering end user is not an importer
Duties of the foreign supplierThose of a manufacturer placing on the marketEconomic operator with distributor-type duties; copy of the declaration of conformity to Swissmedic on request [MedDO, Art. 7(3)]

According to Isler, the legislator itself assumed that a foreign online provider must always appoint a CH-REP; he considers that assumption incorrect (Isler, LSR 1/2024, pp. 22-32). Swiss Medtech discusses Art. 7(1bis) mainly for Swiss online shops and flags the open question whether it covers B2B sales to institutions at all (Swiss Medtech, May 2024, slide 21).

All sources agree on one point: in a direct sale, nobody in Switzerland becomes the importer. That is the setup in the title: an EU manufacturer, possibly with a CH-REP, and no Swiss importer.

Our recommendation: if you actively offer devices to Swiss users, follow the Swissmedic reading. Appoint a CH-REP, put it on the label and register in swissdamed. The cost is small compared with the risk of an authority treating your devices as placed on the market without one.

Why a CH-REP pays off even when it is not compulsory

Even for pure direct use, a CH-REP makes the manufacturer the easier supplier to buy from. Swissmedic advises healthcare institutions to procure, as a rule, from a Swiss manufacturer or one with a CH-REP, and to use devices without a CH-REP only in justified exceptional cases (MU600_00_006, section 7). The reasons it gives are the hospital’s reasons to ask for one:

Liability: the CH-REP is jointly and severally liable with the manufacturer toward an injured party. Devices without a CH-REP may not be covered by Art. 47d TPA.

Safety information: without a Swiss economic operator, Swissmedic is not necessarily informed of field safety corrective actions and cannot necessarily publish them or answer questions.

Responsibility shifts to the hospital: without a CH-REP, the institution must secure the information flow, obtain information, implement corrective actions and clarify liability on its own.

With a CH-REP in place, the formal obligations sit where they belong:

TopicWith CH-REP, direct deliveryLegal basis
Indication on the deviceManufacturer and CH-REP on the label; no importer to indicateMedDO Art. 51; MU600_00_016, section 6
Economic operator registrationCH-REP registers with Swissmedic and receives a CHRNMedDO Art. 55(1) and (4)
Device registration in swissdamedRequired for every device placed on the market since July 1, 2026, including models sold before; due by December 31, 2026MedDO Art. 17(5), Art. 108(2)
Serious incidents and FSCA in SwitzerlandThe CH-REP is responsible for reporting to Swissmedic; the transfer must be agreed in writing in the mandateMedDO Art. 66(1) and (2bis)
Incidents observed by the hospitalReported by the professional to the supplier and to SwissmedicMedDO Art. 66(4)

Takeaways

For an EU manufacturer delivering directly to Swiss healthcare institutions:

Check who places the device on the market. If the hospital buys for its own use, nobody does, and there is no Swiss importer.

If you offer devices to Swiss users online or by other distance selling, assume that Swissmedic sees this as placing on the market, even though legal scholars disagree.

Appoint a CH-REP in either case. Swissmedic requires it for distance selling, and hospitals expect it for direct use.

Do not invent an importer. Without a Swiss entity that places the device on the market, there is none, and none needs to appear on the label.

Tell your customers that passing devices on to another legal entity makes them the importer.

Register your devices in swissdamed through your CH-REP. The obligation applies to every device placed on the Swiss market since July 1, 2026, including models already sold before that date; registration is due by December 31, 2026 [MedDO, Art. 17(5) and 108(2)]. Devices placed on the market only before July 1, 2026, must be registered without delay once a serious incident, FSCA or trend has to be reported [MedDO, Art. 108(3)].

Sources

Swiss Medical Devices Ordinance (MedDO), SR 812.213, as of July 1, 2026

Swissmedic, Information sheet Procurement of medical devices in health institutions, MU600_00_006, version 6.1

Swissmedic, Information sheet Obligations Economic Operators CH, MU600_00_016, version 8.0

Prager Dreifuss for Swiss Medtech, Kurzgutachten Inverkehrbringen, April 22, 2022

Michael Isler, Direkte Beschaffung und Anwendung von Medizinprodukten aus dem Ausland, LSR 1/2024, pp. 22-32

Daniel Delfosse, Swiss Medtech, Medizinprodukteregulierung MDR und MepV, May 2024

Ludger J. Möller, MDSS GmbH, Importer: Myths vs Facts, workshop presentation, December 11, 2024

Decomplix, Swiss medical device importers

Emergo by UL, Medical device direct imports to users in Switzerland (December 2021